Comprehensive Physiology Wiley Online Library

Principles of Measurement: Applications to Pressure, Volume, and Flow

Full Article on Wiley Online Library



Abstract

The sections in this article are:

1 Preliminary Considerations
2 Loading
3 Instrument Response
4 Signal Processing
5 Interpretation of Measurements
5.1 Indirect Measurements (Inference)
5.2 Models
6 Noise
6.1 Intrinsic Noise
6.2 Extrinsic Noise
7 Calibration
7.1 Static Calibrations
7.2 Dynamic Calibrations
7.3 Additional Procedures for Specific Variables
8 Probes
9 Linear Versus Nonlinear Systems
10 Conclusion
Figure 1. Figure 1.

Schematic diagram of the entire measurement process.

Figure 2. Figure 2.

Electrical analogue of the phenomenon of loading. The biological system has a source impedance ZS, whereas the measuring device has an input impedance that is functionally a load ZL. A: pressure or voltage loading. The measured value E1 systematically deviates from the desired value E0 by virtue of a nonzero ZS and a finite ZL. B: flow or current loading. The measured value I1 systematically deviates from the desired value I0 by virtue of a finite ZS and a nonzero ZL.

Figure 3. Figure 3.

Magnitude (A) and phase (B) of the second‐order transfer function T(ω) = [1 + 2iζω/ω0 – (ω/ω0)2]−1 as a function of normalized frequency, as a family in the damping coefficient ζ; ω is the angular frequency 2πf. Dotted line shows the response of a first‐order transfer function T(ω) = (1 + iωτ)−1, where τ is the time constant of the system. Abscissa for the second‐order transfer function is ω/ω0, whereas that for the first‐order is ωτ.

Adapted from Shearer et al. 29
Figure 4. Figure 4.

Effect of connectors on the amplitude and phase response of a Validyne MP‐45 transducer with 2‐cmH2O sensitivity. Unlabeled lines: minimum attachments; Luer: male‐male Luer slip fitting; Stub: needle stub adapter; a, b, c: 15‐cm, 30‐cm, and 45‐cm PE‐200 air‐filled catheters, respectively.

From Jackson and Vinegar 17
Figure 5. Figure 5.

Electrical analogue of a flow plethysmograph. Cg, box gas compliance; R, flow resistance of a pneumotachograph; Pbox, box pressure; , flow rate. Note that ground for resistor is atmospheric pressure, and ground for capacitor is zero pressure. These 2 points nevertheless represent a common virtual ground for the signals of interest.

Figure 6. Figure 6.

Modification of a hypothetical straight‐line flow‐volume curve. A: effect of first‐order transducer system with various time constants (T); τ is the lung time constant for exponential flow decay. B: effect of second‐order transducer system that is slightly underdamped, showing the characteristic “bounce” that accompanies rapid flow onset. Note that flow is shown in units of vital capacities per lung time constant (VC/τ). RV, residual volume; TLC, total lung capacity; ω0, natural frequency; ζ, damping coefficient.

From Sinnett et al. 31
Figure 7. Figure 7.

Magnitude‐frequency plot showing partial compensation of chamber characteristics (C) by judicious choice of transducer characteristics (T). Overall response of the system (S) is remarkably flat.

From Sinnett et al. 31
Figure 8. Figure 8.

Diagram of a dual‐chamber system for volume and flow calibration. Sinusoidally driven loudspeaker separates the reference chamber (volume, V1; pressure, P1) from the testing chamber (volume, V2; pressure, P2).

Adapted from Jackson and Vinegar 17


Figure 1.

Schematic diagram of the entire measurement process.



Figure 2.

Electrical analogue of the phenomenon of loading. The biological system has a source impedance ZS, whereas the measuring device has an input impedance that is functionally a load ZL. A: pressure or voltage loading. The measured value E1 systematically deviates from the desired value E0 by virtue of a nonzero ZS and a finite ZL. B: flow or current loading. The measured value I1 systematically deviates from the desired value I0 by virtue of a finite ZS and a nonzero ZL.



Figure 3.

Magnitude (A) and phase (B) of the second‐order transfer function T(ω) = [1 + 2iζω/ω0 – (ω/ω0)2]−1 as a function of normalized frequency, as a family in the damping coefficient ζ; ω is the angular frequency 2πf. Dotted line shows the response of a first‐order transfer function T(ω) = (1 + iωτ)−1, where τ is the time constant of the system. Abscissa for the second‐order transfer function is ω/ω0, whereas that for the first‐order is ωτ.

Adapted from Shearer et al. 29


Figure 4.

Effect of connectors on the amplitude and phase response of a Validyne MP‐45 transducer with 2‐cmH2O sensitivity. Unlabeled lines: minimum attachments; Luer: male‐male Luer slip fitting; Stub: needle stub adapter; a, b, c: 15‐cm, 30‐cm, and 45‐cm PE‐200 air‐filled catheters, respectively.

From Jackson and Vinegar 17


Figure 5.

Electrical analogue of a flow plethysmograph. Cg, box gas compliance; R, flow resistance of a pneumotachograph; Pbox, box pressure; , flow rate. Note that ground for resistor is atmospheric pressure, and ground for capacitor is zero pressure. These 2 points nevertheless represent a common virtual ground for the signals of interest.



Figure 6.

Modification of a hypothetical straight‐line flow‐volume curve. A: effect of first‐order transducer system with various time constants (T); τ is the lung time constant for exponential flow decay. B: effect of second‐order transducer system that is slightly underdamped, showing the characteristic “bounce” that accompanies rapid flow onset. Note that flow is shown in units of vital capacities per lung time constant (VC/τ). RV, residual volume; TLC, total lung capacity; ω0, natural frequency; ζ, damping coefficient.

From Sinnett et al. 31


Figure 7.

Magnitude‐frequency plot showing partial compensation of chamber characteristics (C) by judicious choice of transducer characteristics (T). Overall response of the system (S) is remarkably flat.

From Sinnett et al. 31


Figure 8.

Diagram of a dual‐chamber system for volume and flow calibration. Sinusoidally driven loudspeaker separates the reference chamber (volume, V1; pressure, P1) from the testing chamber (volume, V2; pressure, P2).

Adapted from Jackson and Vinegar 17
References
 1. Agostoni, E., and G. Miserocchi. Vertical gradient of transpulmonary pressure with active and artificial lung expansion. J. Appl. Physiol. 29: 705–712, 1970.
 2. Bracewell, R. The Fourier Transform and Its Application. New York: McGraw‐Hill, 1965.
 3. Brown, R., F. G. Hoppin, Jr., R. H. Ingram, Jr., N. A. Saunders, and E. R. McFadden, Jr. Influence of abdominal gas on the Boyle's law determination of thoracic gas volume. J. Appl. Physiol.: Respirat. Environ. Exercise Physiol. 44: 469–473, 1978.
 4. Butler, J. P., J. A. Reeds, and S. V. Dawson. Estimating solutions of first kind integral equations with non‐negativity constraints and optimal smoothing. SIAM J. Num. Anal. 18: 381–397, 1981.
 5. D'Angelo, E., M. V. Bonanni, S. Michelini, and E. Agostoni. Topography of the pleural surface pressure in rabbits and dogs. Respir. Physiol. 8: 204–229, 1970.
 6. DuBois, A. B., S. Y. Botelho, and J. H. Comroe, Jr. A new method for measuring airway resistance in man using a body plethysmograph: values in normal subjects and in patients with respiratory disease. J. Clin, Invest. 35: 327–335, 1956.
 7. DuBois, A. B., A. W. Brody, D. H. Lewis, and B. F. Burgess, Jr. Oscillation mechanics of lungs and chest in man. J. Appl. Physiol. 8: 587–594, 1956.
 8. Evans, J. W., and P. D. Wagner. Limits on V.A/Q. distributions from analysis of experimental inert gas elimination. J. Appl. Physiol.: Respirat. Environ. Exercise Physiol. 42: 889–898, 1977.
 9. Finucane, K. E., B. A. Egan, and S. V. Dawson. Linearity and frequency response of pneumotachographs. J. Appl. Physiol. 32: 121–126, 1972.
 10. Fry, D. L. Physiologic recording by modern instruments with particular reference to pressure recording. Physiol. Rev. 40: 753–788, 1960.
 11. Fung, Y. C. A theory of elasticity of the lung. J. Appl. Mech. 41: 8–14, 1974.
 12. Griffin, P. M., and N. Zamel. Volume‐displacement body plethysmograph using a large flowmeter without pressure compensation. J. Appl. Physiol.: Respirat. Environ. Exercise Physiol. 47: 1127–1130, 1979.
 13. Habib, M. P., and L. A. Engel. Influence of the panting technique on the plethysmographic measurement of thoracic gas volume. Am. Rev. Respir. Dis. 117: 265–271, 1978.
 14. Harf, A., G. Atlan, H. Lorino, S. Deshayes, C. Morin, and D. Laurent. Correction for nonlinearity of body flow plethysmograph. J. Appl. Physiol. 50: 658–662, 1981.
 15. Horowitz, P., and W. Hill. The Art of Electronics. Cambridge, UK: Cambridge Univ. Press, 1980.
 16. Jackson, A. C., J. P. Butler, E. J. Millet, F. G. Hoppin, Jr., and S. V. Dawson. Airway geometry by analysis of acoustic pulse response measurements. J. Appl. Physiol.: Respirat. Environ. Exercise Physiol. 43: 523–536, 1977.
 17. Jackson, A. C., and A. Vinegar. A technique for measuring frequency response of pressure, volume, and flow transducers. J. Appl. Physiol.: Respirat. Environ. Exercise Physiol. 47: 462–467, 1979.
 18. Leith, D. E., and J. Mead. Principles of Body Plethysmography. Bethesda, MD: Natl. Heart Lung Inst., Div. Lung Dis., 1974.
 19. Macklem, P. T. Procedures for Standardized Measurements of Lung Mechanics. Bethesda, MD: Natl. Heart Lung Inst., Div. Lung Dis., 1974.
 20. Mathews, J., and R. L. Walker. Mathematical Methods of Physics. New York: Benjamin, 1964.
 21. Mead, J., and J. Milic‐Emili. Theory and methodology in respiratory mechanics with glossary of symbols. In: Handbook of Physiology. Respiration, edited by W. O. Fenn and H. Rahn. Washington, DC: Am. Physiol. Soc, 1964, sect. 3, vol. I, chapt. 11, p. 363–376.
 22. Michaelson, E. D., E. D. Grassman, and W. R. Peters. Pulmonary mechanics by spectral analysis of forced random noise. J. Clin. Invest. 56: 1210–1230, 1975.
 23. Olszowka, A. J. Can V.A/Q. distributions in the lung be recovered from inert gas retention data? Respir. Physiol. 25: 191–198, 1975.
 24. Peslin, R., J. Morinet‐Lambert, C. Duvivier. Étude de la réponse en fréquence de pneumotachographes. Bull. Physio‐Pathol. Respir. 8: 1363–1376, 1972.
 25. Peslin, R., J. Papon, C. Duvivier, and J. Richalet. Frequency response of the chest: modeling and parameter estimation. J. Appl. Physiol. 39: 523–534, 1975.
 26. Pimmel, R. L., M. J. Tsai, D. C. Winter, and P. A. Bromberg. Estimating central and peripheral respiratory resistance. J. Appl. Physiol.: Respirat. Environ. Exercise Physiol. 45: 375–380, 1978.
 27. Rabiner, L. R., and B. Gold. Theory and Application of Digital Signal Processing. Englewood Cliffs, NJ: Prentice‐Hall, 1975.
 28. Roth, P. R. Effective measurements using digital signal analysis. IEEE Spectrum 8: 62–70, 1971.
 29. Shearer, J. L., A. T. Murphy, and H. H. Richardson. Introduction to Systems Dynamics. Reading, MA: Addison‐Wesley, 1967.
 30. Sidell, R. S., and J. J. Fredberg. Noninvasive inference of airway network geometry from broad band lung reflection data. J. Biomech. Eng. 100: 131–138, 1978.
 31. Sinnett, E. E., A. C. Jackson, D. E. Leith, and J. P. Butler. Fast integrated flow plethysmograph for small mammals. J. Appl. Physiol.: Respirat. Environ. Exercise Physiol. 50: 1104–1110, 1981.
 32. Wagner, P. D. Information content of the multibreath nitrogen washout. J. Appl. Physiol.: Respirat. Environ. Exercise Physiol. 46: 579–587, 1979.
 33. Wagner, P. D., H. A. Saltzman, and J. B. West. Measurement of continuous distributions of ventilation‐perfusion ratios: theory. J. Appl. Physiol. 36: 588–599, 1974.
 34. West, J. B., and F. L. Matthews. Stresses, strains, and surface pressures in the lung caused by its weight. J. Appl. Physiol. 32: 332–345, 1972.

Contact Editor

Submit a note to the editor about this article by filling in the form below.

* Required Field

How to Cite

James P. Butler, David E. Leith, Andrew C. Jackson. Principles of Measurement: Applications to Pressure, Volume, and Flow. Compr Physiol 2011, Supplement 12: Handbook of Physiology, The Respiratory System, Mechanics of Breathing: 15-33. First published in print 1986. doi: 10.1002/cphy.cp030302